electron multiplier

Also found in: Dictionary, Financial, Encyclopedia, Wikipedia.

electron multiplier

A detection device inside the vacuum of a mass spectrometer which converts ions impacting on its front dynodes into a detectable, amplified electron current at the back lead of the device, with an overall gain (signal out/signal in) as high as 104–108.

Positive ions exiting the mass analyser impact the first dynode’s surface, triggering the release of several electrons, which are then accelerated through a potential to the next electrode, where each electron impact triggers the release of several secondary electrons; these are accelerated into the next dynode, leading to further impact (release events). A cascade of electrons is produced, generating a current that is further amplified and then sampled by an analog-to-digital converter to be recorded by the data system.
Segen's Medical Dictionary. © 2012 Farlex, Inc. All rights reserved.
Mentioned in
References in periodicals archive
During the RF sweep, which progressively destabilizes ions for ejection from the ion trap to the electron multiplier, the axial modulation RF enhances the movement of ions away from the center of the ion trap where magnetic field strength is low and ion density is high.
Additional studies done by diluting the low-concentration control demonstrated accurate quantification at 1 [micro]g/L (electron multiplier set at 2000 eV, injection volume = 2 [micro]L).
Copyright © 2003-2025 Farlex, Inc Disclaimer
All content on this website, including dictionary, thesaurus, literature, geography, and other reference data is for informational purposes only. This information should not be considered complete, up to date, and is not intended to be used in place of a visit, consultation, or advice of a legal, medical, or any other professional.